Radionuclide Information
NUCLIDE: H-3
TYPE(S) OF RADIATION |
ENERGY |
PERCENT DISINTEGRATION |
RANGE IN AIR |
RANGE IN WATER |
SHIELDING REQUIRED |
|
|
|
|
|
|
Beta |
18.6 keV (max) |
100% |
6 mm |
0.006 mm |
None |
Physical half-life 12.3 years
Survey instruments: none practical
Removable quantification
method: smear and count in liquid scintillation counter
Personal dosimeter: none required
NUCLIDE: C-14
TYPE(S) OF RADIATION |
ENERGY |
PERCENT DISINTEGRATION |
RANGE IN AIR |
RANGE IN WATER |
SHIELDING REQUIRED |
|
|
|
|
|
|
Beta |
156 keV (max) |
100% |
24 cm |
0.28 mm |
Plexiglass shielding reduces dose, but
for microcurie amounts none is required |
Physical half-life: 5730 years
Survey instruments:
GM use for gross contamination is possible, but efficiency is
poor.
Removable quantification
method: smear and count in liquid scintillation counter
Personal dosimeter:
none required
NUCLIDE: S-35
TYPE(S) OF RADIATION |
ENERGY |
PERCENT DISINTEGRATION |
RANGE IN AIR |
RANGE IN WATER |
SHIELDING REQUIRED |
|
|
|
|
|
|
Beta |
167 keV (max) |
100% |
26 cm |
0.32 mm |
Plexiglass shielding reduces dose, but
for microcurie amounts none is required |
Physical half-life:
87.4 days
Survey instruments:
GM use for gross contamination is possible, but efficiency is
poor.
Removable quantification
method: smear and count in liquid scintillation counter
Personal dosimeter:
none required
NUCLIDE: P-32
TYPE(S) OF RADIATION |
ENERGY |
PERCENT DISINTEGRATION |
RANGE IN AIR |
RANGE IN WATER |
SHIELDING REQUIRED |
|
|
|
|
|
|
| Beta |
1.71 MeV (max) |
100% |
790 cm |
0.76 cm |
1 cm Plexiglass |
Physical half-life: 14.3 days
Survey instruments:
GM survey meter
Removable quantification
method: smear and count in liquid scintillation counter
Personal dosimeter:
Whole body and extremity dosimeters
NUCLIDE: P-33
TYPE(S) OF RADIATION |
ENERGY |
PERCENT DISINTEGRATION |
RANGE IN AIR |
RANGE IN WATER |
SHIELDING REQUIRED |
|
|
|
|
|
|
| Beta |
249 keV (max) |
100% |
49 cm |
0.6 mm |
Plexiglass shielding reduces dose, but
for microcurie amounts none is required |
Physical half-life: 25.4days
Survey instruments:
GM survey meter
Removable quantification
method: smear and count in liquid scintillation counter
Personal dosimeter:
Whole body and extremity dosimeters for higher activities
NUCLIDE: I-125
TYPE(S) OF RADIATION |
ENERGY |
PERCENT DISINTEGRATION |
SHIELDING REQUIRED |
|
|
|
|
Te K X-rays |
27-32 keV |
140 % |
Half-value-layer is 0.02 mm Pb |
| |
35 keV |
7% emitted;
93% internally
converted
|
|
Physical half-life: 60.2 days
Volatility (@STP): substantial in disassociated form or in
acidic solutions
Survey instruments:
thin-window NaI scintillator; GM survey meter is less efficient.
Removable quantification
method: smear and count in liquid scintillation counter or gamma
counter
Personal dosimeter:
Whole body and extremity dosimeters for higher activities
|